edoc-Server der Humboldt-Universität zu Berlin

Dissertation

Autor(en): Seyed Madjid Hadjizadeh-Ziabari
Titel: Subjektive Bildqualität digitaler Panoramaschichtaufnahmen in Relation zur Exposition
Gutachter: F. Sitzmann; Peter A. Reichart; W. Hopfenmüller
Erscheinungsdatum: 15.07.2002
Volltext: pdf (urn:nbn:de:kobv:11-10018235)
Fachgebiet(e): Medizin
Schlagwörter (ger): Bildqualität, Panoramaaufnahme, Strahlenschutz, Röntgenstrahlen
Schlagwörter (eng): radiography, digital panoramic, exposure, image quality
Einrichtung: Humboldt-Universität zu Berlin, Medizinische Fakultät - Universitätsklinikum Charité
Zitationshinweis: Hadjizadeh-Ziabari, Seyed Madjid: Subjektive Bildqualität digitaler Panoramaschichtaufnahmen in Relation zur Exposition; Dissertation, Humboldt-Universität zu Berlin, Medizinische Fakultät - Universitätsklinikum Charité , publiziert am 15.07.2002, urn:nbn:de:kobv:11-10018235
Metadatenexport: Um den gesamten Metadatensatz im Endnote- oder Bibtex-Format zu speichern, klicken Sie bitte auf den entsprechenden Link. Endnote   Bibtex  
print on demand: Wenn Sie auf dieses Icon klicken, können Sie ein Druckexemplar dieser Publikation bestellen.
Diese Seite taggen: Diese Icons führen auf so genannte Social-Bookmark-Systeme, auf denen Sie Lesezeichen anlegen, persönliche Tags vergeben und Lesezeichen anderer Nutzer ansehen können.
  • connotea
  • del.icio.us
  • Furl
  • RawSugar

Abstract (ger):
Die Zielsetzung dieser Studie war es, für digitale Panoramaschichtaufnahmen die Relation der subjektiven Bildqualität zur Exposition zu bestimmen. An Hand von Humanpräparataufnahmen konnten bei Standardwerten sowie experimentellen Variationen der Expositionsdaten weiterhin Effekte der intentionellen Unterexposition sowie der Strahlenaufhärtung auf die subjektive Bildqualität quantifiziert werden. Die Herstellung der Aufnahmen erfolgte auf einem Sirona Orthophos DS. Dabei wurden 37 Aufnahmen mit Expositionswerten von 60 kV/9 mA - 84 kV/13 mA aus dem Herstellerprogramm erzeugt und mit Hilfe einer modifizierten Steuerungssoftware zusätzliche PSA mit experimentellen Einstellungen von 60 kV/3 mA - 90 kV/11 mA hergestellt. Für die Beurteilung und Auswertung der subjektiven Bildqualität wurde eine individuelle Software (Eldoredo V2.2) entwickelt. 39 Zahnärzte und 5 MTRA beurteilten die Aufnahmen damit an einem Monitor unter standardisierten Bedingungen. Ein iterativer Beurteilungsprozess erlaubte, eine Serie von 1369 (37 x 37) PSA-Abbildungspaaren darzustellen. Für jedes Paar entschieden die Untersucher an Hand definierter Kriterien, ob eine PSA hinsichtlich der subjektiv beurteilten Bildqualität vorzuziehen oder Äquivalenz gegeben sei. Nach statistischer Aufarbeitung der Einzelentscheidungen ließ sich damit für jede Expositionsstufe ein Index der Bildqualität berechnen. Bei Expositionswerten in einem Bereich von 60 kV/9 mA - 69 kV/15 mA der Herstellersoftware und 60 kV/5 - 15 mA sowie 70 kV/5 - 15mA der experimentellen Software fanden sich dabei keine signifikanten Verteilungsunterschiede der Bildqualität. Eine intentionelle Unterexposition bei digitalen PSA-Geräten, etwa bei Kindern oder häufigen Wiederholungsaufnahmen, kann nach den vorliegenden Ergebnissen vertreten werden, ohne dass es dabei zu einer signifikanten Verschlechterung der Bildqualität kommt. Damit ist im Gegensatz dazu bei einer Strahlenaufhärtung in dem untersuchten digitalen System stets zu rechnen. Insgesamt zeigen die Ergebnisse, dass auch digitale PSA-Systeme beachtliche Reserven hinsichtlich der Dosisminimierung aufweisen können.
Abstract (eng):
The aim of this study was to describe the relation of the subjective image quality of digital panoramic radiographs in relation to exposure. In addition, variations of exposure were compared to standard settings, thus evaluating the effects of intentional underexposure on the achievable image quality. A Sirona Orthophos DS unit was used to produce 37 digital panoramic images of a human skull. Exposure values ranged from 60 kV/9 mA to 84 kV/13 mA in the conventional and 60 kV/3 mA to 90 kV/11 mA in the experimental setting. Assessment and evaluation of the subjective image quality were performed with an HTML-based protocol. 39 dentists and 5 radiographic assistants had to assign their preference of an image or equality in 1,369 (37x37) image pairs. The decisions were computed to a quality index for each exposure setting. Statistical analysis demonstrated no significant differences of image quality between 60 kV/9 mA - 69 kV/15 mA in the conventional and 60 kV/5 to15 mA as well as 70 kV/5 to 15 mA in the experimental setting. Following these results, a considerable dose reduction by the means of intentional underexposure can be achieved without any loss of image quality. By reducing the absorbed doses, an increase of kV values up to 80 kV and more is also correlated with dose reduction. However, those images showed high significant loss of quality. In summary, the results demonstrate an equivalent image quality of digital panoramic images over a very wide range of exposure values. The feasible dose reduction might be of interest not only in individuals (minors, repeated images), but also in defining general principles of panoramic imaging.
Zugriffsstatistik: Die Daten für die Zugriffsstatistik der einzelnen Dokumente wurden aus den durch AWStats aggregierten Webserver-Logs erstellt. Sie beziehen sich auf den monatlichen Zugriff auf den Volltext sowie auf die Startseite. Die Zugriffsstatistik wird nicht standardisiert erfasst und kann maschinelle Zugriffe enthalten.
 
Bei Formatversionen eines Dokuments, die aus mehreren Dateien bestehen (insbesondere HTML), wird jeweils der monatlich höchste Zugriffswert auf eine der Dateien (Kapitel) des Dokuments angezeigt.
 
Um die detaillierten Zugriffszahlen zu sehen, fahren Sie bitte mit dem Mauszeiger über die einzelnen Balken des Diagramms.
Startseite: 2 Zugriffe PDF: 12 Zugriffe PDF: 7 Zugriffe Startseite: 5 Zugriffe PDF: 15 Zugriffe Startseite: 3 Zugriffe PDF: 29 Zugriffe Startseite: 2 Zugriffe PDF: 21 Zugriffe Startseite: 2 Zugriffe PDF: 15 Zugriffe PDF: 23 Zugriffe PDF: 18 Zugriffe PDF: 28 Zugriffe Startseite: 3 Zugriffe PDF: 34 Zugriffe Startseite: 4 Zugriffe PDF: 29 Zugriffe PDF: 16 Zugriffe PDF: 13 Zugriffe PDF: 13 Zugriffe PDF: 22 Zugriffe PDF: 12 Zugriffe Startseite: 3 Zugriffe PDF: 44 Zugriffe Startseite: 3 Zugriffe PDF: 40 Zugriffe Startseite: 2 Zugriffe PDF: 17 Zugriffe Startseite: 3 Zugriffe PDF: 26 Zugriffe Startseite: 1 Zugriffe PDF: 22 Zugriffe Startseite: 2 Zugriffe PDF: 24 Zugriffe Startseite: 2 Zugriffe PDF: 25 Zugriffe PDF: 28 Zugriffe Startseite: 1 Zugriffe PDF: 10 Zugriffe Startseite: 4 Zugriffe PDF: 36 Zugriffe Startseite: 4 Zugriffe PDF: 23 Zugriffe Startseite: 2 Zugriffe PDF: 34 Zugriffe Startseite: 4 Zugriffe PDF: 31 Zugriffe PDF: 25 Zugriffe Startseite: 1 Zugriffe PDF: 22 Zugriffe Startseite: 2 Zugriffe PDF: 25 Zugriffe Startseite: 3 Zugriffe PDF: 30 Zugriffe PDF: 31 Zugriffe Startseite: 2 Zugriffe PDF: 45 Zugriffe Startseite: 2 Zugriffe PDF: 38 Zugriffe PDF: 41 Zugriffe
Jul
11
Aug
11
Sep
11
Oct
11
Nov
11
Dec
11
Feb
12
Apr
12
May
12
Jun
12
Jul
12
Aug
12
Sep
12
Oct
12
Nov
12
Dec
12
Jan
13
Feb
13
Mar
13
Apr
13
May
13
Jun
13
Jul
13
Aug
13
Sep
13
Oct
13
Nov
13
Dec
13
Jan
14
Feb
14
Mar
14
Apr
14
May
14
Jun
14
Jul
14
Aug
14
Sep
14
Monat Jul
11
Aug
11
Sep
11
Oct
11
Nov
11
Dec
11
Feb
12
Apr
12
May
12
Jun
12
Jul
12
Aug
12
Sep
12
Oct
12
Nov
12
Dec
12
Jan
13
Feb
13
Mar
13
Apr
13
May
13
Jun
13
Jul
13
Aug
13
Sep
13
Oct
13
Nov
13
Dec
13
Jan
14
Feb
14
Mar
14
Apr
14
May
14
Jun
14
Jul
14
Aug
14
Sep
14
Startseite 2   5 3 2 2       3 4           3 3 2 3 1 2 2   1 4 4 2 4   1 2 3   2 2  
PDF 12 7 15 29 21 15 23 18 28 34 29 16 13 13 22 12 44 40 17 26 22 24 25 28 10 36 23 34 31 25 22 25 30 31 45 38 41

Gesamtzahl der Zugriffe seit Jul 2011:

  • Startseite – 62 (1.68 pro Monat)
  • PDF – 924 (24.97 pro Monat)
 
 
Generiert am 24.10.2014, 03:38:39